Available Sensor Options:
| Range | Resolution | Accuracy | |
| Temperature (°C, °F, K) | -5 to 55 °C | 0.1 °C | ±0.3 °C |
| Dissolved Oxygen % Saturation (galvanic or polarographic) | 0 to 500% air saturation | 0.1% or 1%% air saturation (selectable) |
0 to 200%: ±2% of the reading or ±2% air saturation, whichever is greater; 200 to 500%: ±6% of the reading |
| Dissolved Oxygen mg/L (galvanic or polarographic) | 0 to 50 mg/L | 0.01 or 0.1 mg/L (selectable) |
0 to 20 mg/L: ±2% of the reading or ±0.2mg/L, whichever is greater; 20 to 50 mg/L: ±6% of the reading |
| Barometer (mmHg, inHg, mbar, psi, kPa) | 500 to 800 mm Hg | 0.1 mm Hg | ±3 mm Hg within ±15 ºC range from calibration point |