Horiba TB220 Turbidity Calibration Standard Solutions
Sealed calibration standards for the HORIBA LAQUA TB220 portable turbidity meter. Each standard is supplied in a 50 mL vial at a certified NTU value, ready to place straight into the instrument — no dilution, no formazin preparation, no handling of hazardous stock suspensions.
Choose your set
- 550-TBP-4 (part no. 3201063378) — full four-point set: 0.02, 20, 100 and 800 NTU. Calibrates the meter across its entire measuring range and is the right choice for a new instrument or an annual full recalibration.
- 550-TBP-2L (part no. 3201063380) — low set: 0.02 and 20 NTU. For work concentrated at the clean end, such as finished drinking water and filter effluent.
- 550-TBP-2H (part no. 3201063381) — high set: 100 and 800 NTU. For turbid work such as raw water, stormwater and sediment monitoring.
Why the 0.02 NTU standard matters
The low-end standard is what makes trustworthy drinking water measurement possible. Regulatory turbidity limits for filtered water sit well below 1 NTU, so an instrument calibrated only at higher values is being asked to extrapolate into precisely the range where the decision gets made. Calibrating at 0.02 NTU anchors the meter where your compliance readings actually fall.
Handling for accurate results
- Invert sealed standards gently before use — never shake, which introduces bubbles that scatter light and inflate readings
- Let the vial come to ambient temperature to prevent condensation on the glass
- Wipe the vial clean and apply silicone oil to mask fine scratches before measuring
- Always insert the vial in the same orientation, using the index mark
- Store out of direct sunlight and replace standards at the end of their stated shelf life
Applications
- Drinking water treatment and distribution compliance calibration
- Filter performance verification and breakthrough monitoring
- Stormwater and construction site discharge monitoring programs
- Surface water and watershed survey work
- Wastewater effluent clarity verification
- Documented instrument calibration under a laboratory quality system