Horiba LAQUAtwin Standard & Calibration Solutions
Genuine HORIBA standard and calibration solutions for LAQUAtwin pocket meters. Most are supplied as sets of six 14 mL sachets — a single-use format that keeps every calibration traceable to fresh, uncontaminated standard. Once a sachet is opened it is used and discarded, so there is no bottle sitting on a shelf slowly absorbing carbon dioxide or growing biofilm.
Calibration is only as good as the standard behind it. Standards drift with age and contaminate quickly once opened, and a meter calibrated against a degraded standard reads confidently wrong. The sachet format removes that failure mode entirely, which is why it matters for any measurement that has to stand up to scrutiny.
pH and conductivity
- 514-4 (3999960108) — pH 4 standard, 6 × 14 mL
- 514-7 (3999960109) — pH 7 standard, 6 × 14 mL
- 514-22 (3999960110) — conductivity 1.41 mS/cm, 6 × 14 mL
- 514-23 (3999960111) — conductivity 12.9 mS/cm, 6 × 14 mL
- 514-20 (3999960114) — moistening solution, for sensor storage and conditioning
Salt and ORP
- 514-05 (3999960112) — salt 5 ppt / 0.5%, 6 × 14 mL
- 514-50 (3999960113) — salt 50 ppt / 5.0%, 6 × 14 mL
- 514-ORP-225 (3200997084) — ORP 255 mV standard, 6 × 14 mL
Nitrate
- Y044 (3200053535) — nitrate 30 ppm, 6 × 14 mL
- Y045 (3200053536) — nitrate 150 ppm, 6 × 14 mL
- Y042 (3200053514) — nitrate 300 ppm, 6 × 14 mL
- Y043 (3200053532) — nitrate 2000 ppm, 6 × 14 mL
- Y041 (3200053433) — nitrate 5000 ppm, 6 × 14 mL
Sodium, potassium and calcium
- Y022L (3200457724) — sodium 150 ppm, 6 × 14 mL
- Y022H (3200457723) — sodium 2000 ppm, 6 × 14 mL
- Y031L (3200457720) — potassium 150 ppm, 6 × 14 mL
- Y031H (3200457719) — potassium 2000 ppm, 6 × 14 mL
- Y051L (3200457728) — calcium 150 ppm, 6 × 14 mL
- Y051H (3200457727) — calcium 2000 ppm, 6 × 14 mL
Fluoride
- 514-F-1 (3200991628) — fluoride 1 ppm, 6 × 14 mL
- 514-F-10 (3200991630) — fluoride 10 ppm, 6 × 14 mL
- 514-F-TISAB (3200991632) — total ionic strength adjustment buffer, 6 × 14 mL. Required for accurate fluoride measurement.
Ca-11C blood calcium consumables
- Y053 (3200774921) — complete standard solution kit: 2 × Y052L, 2 × Y052H and 3 × #251 cleaning solution
- Y052L (3200774598) — 1.25 mmol/L standard, 2 × 25 mL bottles
- Y052H (3200774599) — 2.50 mmol/L standard, 2 × 25 mL bottles
- #251 (3200774601) — cleaning solution, 250 mL bottle
Calibration guidance
- Calibrate with two standards that bracket your expected sample range for best accuracy.
- Recalibrate after fitting a new sensor, and routinely at the start of each measuring session.
- Use the high standard for ion-selective sensor conditioning before first use.
- Fluoride measurements require TISAB to control ionic strength and remove interference.
Documentation
Replacement sensors, sampling sheets and carrying cases are in the Horiba Parts & Accessories collection. Complete meters are in the HORIBA collection.