Horiba U-52 Specifications

Horiba U-52 Water Quality Meter Sensors

Technical Specifications

 

 Sensor Type Measurement Principle

Range

Resolution

Repeatability

Accuracy

pH

  • 2-point Calibration
  • Automatic Temperature
Glass electrode Method

pH 0 to 14

0.01pH

±0.05pH

± 0.1pH

Oxidation Reduction Potential (ORP) Platinum electrode method

-2000 mV to +2000 mV

1mV

 

±5mV

 

±15mV

Dissolved Oxygen (DO)

  • Salinity conversion (0 to 70 PPT/automatic)
  • Automatic temperature compensation
Polarographic Method

0 to 50.0 mg/L

0.01 mg/L

±0.1 mg/L

0 to 20 mg/L: ±0.2 mg/L

20 to 50 mg/L: ±0.5mg/L

 

Conductivity (COND)

  • Auto range
  • Automatic Temperature Conversion (25°C)
4 AC electrode Method

0 to 10 S/m (0 to 100 mS/cm)

0.000 to 0.999 mS/cm: 0.001 mS/cm

1.00 to 9.99 mS/cm: 0.01 mS/cm

10.0 to 99.9 mS/cm: 0.1 mS/cm

0.0 to 99.9 mS/m: 0.1 mS/m

0.100 to 0.999 S/m: 0.001 S/m

1.00 to 9.99 S/m: 0.01 S/m

±0.5% F.S.

*±1% F.S (Median of two-point calibration)

Salinity

Conductivity conversion

0 to 70 PPT (permillage)

0.1 PPT

±1 PPT

±3 PPT

Total Dissolved Solid (TDS)

  • Conversion factor Setting
Conductivity conversion

0. too g/L

0.1% F.S.

±2 g/L

±5 g/L

Seawater Specific Gravity

Conductivity Conversion

0 to 50 t

0.1 t

±2 t

±5 t

Temperature

Thermistor method

-10 to 55°C

0.01°C

*±0.10°C ( at calibration point)

JIS Class B Platinum Thermometer sensor (±0.2+0.005 | t |)

Turbidity (TURB)

LED light source and 30° scattering method

0 to 800 NTU

0 to 99.9 NTU: 0.1 NTU

100 to 800 NTU: 1NTU

*±5% (Reading) or ±0.05 NTU whichever is greater

±5% (Reading) or ±1 NTU whichever is greater

Water Depth (U-52 GPS model only)

Pressure Method

0 to 30 m

0.05 m

±1% F.S.

±0.3m

GPS

US-50 Series G Models Only





*Battery life based on continuous operation using alkaline C dry batteries when the monitor temperature is over 20°C and the backlight OFF.
*Accuracy is measure by calibrating 4 points for turbidity and electrical conductivity and 2 points for all other measurements against standard solution.
*Repeatability is measured by the ability to reproduce the results against the standard solution ( at 25°C normal pressure condition)